NO VERSI ASLI VERSI TERJEMAHAN 94 Both temperature and pressure should be used to monitor the process. Control instrumentation should normally be independent of monitoring instrumentation and recording charts. Where automated control and monitoring systems are used for these applications they should be validated to ensure that critical process requirements are met. System and cycle faults should be registered by the system and observed by the operator. The reading of the independent temperature indicator should be routinely checked against the chart recorder during the sterilisation period. For sterilisers fitted with a drain at the bottom of the chamber, it may also be necessary to record the temperature at this position, throughout the sterilisation period. There should be frequent leak tests on the chamber when a vacuum phase is part of the cycle.
Suhu dan tekanan harus digunakan untuk memantau proses sterilisasi. Instrumen pengendali biasanya harus independen terhadap instrumen pemantau dan grafik pencatat. Penggunaan sistem pengendali dan pemantau otomatis harus divalidasi untuk memastikan bahwa persyaratan proses kritis terpenuhi. Kesalahan pada sistem dan siklus harus dicatat dalam sistem dan diamati oleh operator. Pembacaan indikator suhu independen harus diperiksa secara rutin dan dibandingkan dengan pencatat grafik selama periode sterilisasi. Untuk sterilisator yang dilengkapi dengan drainase di bagian bawah chamber, mungkin perlu juga dilakukan pencatatan suhu pada posisi ini, sepanjang periode sterilisasi. Harus ada uji kebocoran yang sering dilakukan pada chamber saat fase vakum merupakan bagian dari siklus sterilisasi.
…
Silahkan Daftar atau Login terlebih dahulu untuk mengakses artikel secara keseluruhan.
Register User Login